DELMER LD-Series Analytical Balances
Laboratory Weighing with the DELMER LD-Series
DELMER LD-Series analytical balances are intended for laboratory tasks where small mass differences matter, including analytical sample preparation, reference materials, formulation, research and quality control. The models listed here span capacities from 64 g to 314 g. Select the instrument by capacity, readability and demonstrated performance against your method’s tolerance, not by display resolution alone.

LD-Series Analytical Balance Specifications
The following values reflect the existing DELMER listings. Confirm the current model datasheet, test conditions and configuration before purchase or use in a controlled procedure.
| Model | Capacity | Readability | Repeatability, listed | Linearity, listed | Pan diameter |
|---|---|---|---|---|---|
| LD 64C | 64 g | 0.1 mg | ±0.1 mg | ±0.2 mg | 90 mm |
| LD 124C | 124 g | 0.1 mg | ±0.1 mg | ±0.2 mg | 90 mm |
| LD 224C | 224 g | 0.1 mg | ±0.1 mg | ±0.2 mg | 90 mm |
| LD 314C | 314 g | 0.1 mg | ±0.1 mg | ±0.2 mg | 90 mm |
| LD 285C | 285 g | 0.1 mg / 0.01 mg, listed variant | ±0.1 mg | ±0.2 mg | 70 mm |
The LD 285C listing includes two readability values but does not define their applicable load ranges. Confirm the exact variant and range boundaries before assuming 0.01 mg readability across the full 285 g capacity. Readability is the display increment, not a guarantee of accuracy or measurement uncertainty.
For additional range information, review the DELMER analytical balance model overview.
Laboratory Applications
Analytical Chemistry and Sample Preparation
Weigh reagents, reference materials and research samples when the selected balance’s verified performance meets the procedure’s requirements.
Formulation and Quality Control
Compare capacity, repeatability and required tolerance for small batches, routine laboratory checks and formulation work. Include the container in your total-load assessment.
Research and Materials Testing
Assess the smallest sample, environmental stability and handling method. Confirm any specialised accessories before selecting a balance for filter or other small-sample work.
Finer Low-Mass Weighing
If your procedure requires a finer display increment, compare the DELMER LD 2 and LD 5 microbalances. Their lower capacities and different listed performance need to be evaluated against the actual method.
Calibration, Display and Connectivity
The range listings describe internal calibration, a backlit LCD display and RS-232C serial connectivity, with USB listed as optional. Confirm availability and inclusion on the exact model quotation.
Before ordering, confirm calibration functions and triggers, compatible cables, software requirements and the intended data-capture workflow. Do not assume direct Windows or Microsoft Office transfer from a cable connection alone.
Model-Specific Features to Verify
The previous page also described motorised infrared-controlled doors, below-balance weighing, a specialised filter draft shield, MONO-TECH sensor technology and statistical calculations. These functions need confirmation against the exact model datasheet; they are not presented here as universal standard features.
If statistics are required, specify the records and calculations needed, such as reading counts, minimum and maximum values, standard deviation or coefficient of variation. Confirm what the instrument or accompanying software supports.
How to Choose and Install an Analytical Balance
- Define the load: establish the smallest sample and the maximum combined mass of sample and vessel.
- Set the tolerance: identify allowable error and uncertainty rather than selecting on readability alone.
- Review performance: compare listed repeatability and linearity, and confirm test conditions and minimum usable sample mass.
- Prepare the location: use a rigid, level, low-vibration bench away from airflow, rapid temperature changes and static sources.
- Plan calibration and checks: follow the manufacturer’s instructions and your laboratory’s quality procedures.
- Confirm documentation and data handling: specify required records, interfaces, software and accessories before purchase.
Allow samples and containers to equilibrate to the weighing environment and use consistent handling. A draft shield and internal adjustment do not eliminate every environmental effect or replace performance checks.
Frequently Asked Questions
What does 0.1 mg readability mean?
The display changes in 0.1 mg increments. This does not by itself establish accuracy, uncertainty or the smallest sample your method can reliably weigh.
Which LD-Series model should I choose?
Choose against the full sample-and-container load, smallest sample, required tolerance, pan dimensions and verified performance. The listed models provide different capacities; higher capacity alone does not make a model more suitable.
Does the LD 285C provide 0.01 mg readability at every load?
The existing listing states 0.1 mg / 0.01 mg but does not define the applicable load ranges. Confirm the exact variant and range boundaries in the current datasheet.
Are these balances automatically GLP or GMP compliant?
A product description does not establish regulatory compliance. Suitability depends on the selected instrument, verified performance, qualification, calibration, documentation and the procedures required by the laboratory’s quality system.
Are motorised doors and USB included?
Do not assume so. Confirm the draft-shield configuration and interfaces on the exact model. USB is described as optional in the range listing.
When should I compare a microbalance?
Compare the microbalance selection guide when your procedure requires finer low-mass weighing. Check capacity and demonstrated performance as well as readability.
Request Analytical Balance Selection Advice
Share your smallest sample, maximum load including the container, required tolerance, application, destination country and interface needs. Request the current model datasheet and included configuration.
Contact DELMER for Balance Advice